ETHICON ENDO-SURGERY, LLC. OPT SLEEVE 5X100MM STABILITY; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY Back to Search Results
Model Number 2CB5LT
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported during a cholecystectomy the ethicon xcel 5mm trocar was "sticking".There were no patient consequences.
 
Additional Manufacturer Narrative
(b)4).A photo was submitted to ethicon endo surgery for review along with the device sample.During the visual analysis of the photo, the following was observed: the photo shows a trocar with a broken sleeve, with a device inserted.Visual analysis of the returned sample determined that the 2cb5lt device was received with the tip of the sleeve damaged melted.No sticking was observed during the analysis.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.As part of ethicon¿s quality process all devices are manufactured, inspected, and released to approved specifications.In order to prevent this kind of damage please avoid the contact with laser, electrosurgical or ultrasonic devices.No further investigation will be conducted on this complaint due to external cause.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPT SLEEVE 5X100MM STABILITY
Common Device Name
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo
*  
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
*  
MDR Report Key14013837
Report Number3005075853-2022-02055
Device Sequence Number20008210
Product Code GCJ
UDI-Device Identifier10705036013730
UDI-Public10705036013730
Combination Product (Y/N)N
PMA/510(K) Number
K032676
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2026
Device Model Number2CB5LT
Device Catalogue Number2CB5LT
Device Lot Number409A59
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/09/2022
Initial Report FDA Received Date04/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-