• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ADVANIX? PANCREATIC STENT; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ADVANIX? PANCREATIC STENT; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00536540
Medical Device Problem Codes Break (1069); Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a previously implanted advanix pancreatic stent was being removed from a patient during an endoscopic pancreatic duct stent replacement procedure from the bile duct performed on (b)(6) 2022.During the planned stent replacement procedure, a pair of grasping forceps was used to remove the previously implanted stent, however, it became separated.It was reported that the broken stent was successfully removed from the patient and another advanix pancreatic stent was implanted and successfully completed the procedure.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.Initial reporter state: (b)(6).(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a problem analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a previously implanted advanix pancreatic stent was being removed from a patient during an endoscopic pancreatic duct stent replacement procedure from the bile duct performed on (b)(6) 2022.During the planned stent replacement procedure, a pair of grasping forceps was used to remove the previously implanted stent, however, it became separated.It was reported that the broken stent was successfully removed from the patient and another advanix pancreatic stent was implanted and successfully completed the procedure.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
The complainant was unable to provide the suspect device lot number; therefore, the lot expiration and device manufacture dates are unknown.Medical device code a0401 captures the reportable event of stent broken.Conclusion code d17 is being used in lieu of an adequate conclusion code for device not returned.The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a problem analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANIX? PANCREATIC STENT
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, MA 01752
9145716918
MDR Report Key14014053
Report Number3005099803-2022-01536
Device Sequence Number19973220
Product Code FGE
UDI-Device Identifier08714729857488
UDI-Public08714729857488
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K133700
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00536540
Device Catalogue NumberM00536540
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/08/2022
Supplement Date Received by Manufacturer03/25/2026
Initial Report FDA Received Date04/05/2022
Supplement Report FDA Received Date04/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-