| Model Number |
101/858/090 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Device Embedded In Tissue or Plaque (3165); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/08/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the rn removed the inner cannula to exchange a new one at shift changes.Part of inner cannula broke while rn removed it so the inner canular was stuck , rn was unable to remove the broken inner cannula from patient tracheostomy.Th issue was resolved by replacing the inner cannula safely.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One device was received for evaluation in relation to the reported event.Evaluation of the device is currently in progress.Once the evaluation is complete, a follow-up report with evaluation results will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).A product was received for evaluation.The upper ring is broken, thus failure mode is confirmed; however, despite able to confirm the failure mode reported, it is not possible relate defect with process manufacturing.The root cause of the complaint was not able to be determined.No corrective actions were taken since the complaint was not relate to manufacturing process.
|
| |
|
Event or Problem Description
|
|
Additional information indicated that a medical intervention was required to replace the new inner cannula safely.
|
| |
|