DORNOCH DUO FLUID CART; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED Back to Search Results
Model Number N/A
Medical Device Problem Code Smoking (1585)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this cart had smoke coming from it prior to surgery.No adverse events were reported as a result of this malfunction.
 
Additional Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Once an evaluation of this device is completed, a follow-up/final report will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.-review of the most recent repair record determined the unit was confirmed to have a burnt/charred power inlet module.The power inlet module was replaced and resolved the reported issue.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUO FLUID CART
Common Device Name
APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
Manufacturer (Section D)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer (Section G)
DORNOCH
200 northwest parkway
riverside MO 64150
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key14014211
Report Number0001954182-2022-00021
Device Sequence Number20407780
Product Code JCX
UDI-Device Identifier00889024465992
UDI-Public(01)00889024465992
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K172481
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 05/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00514010100
Device Lot Number0027501
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/04/2022
Supplement Date Received by Manufacturer05/02/2022
Initial Report FDA Received Date04/05/2022
Supplement Report FDA Received Date05/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/25/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-