| Model Number |
N/A |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Pain (1994); Ambulation or Postural Difficulties (2544)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported patient underwent right total knee arthroplasty.Subsequently, patient reports pain, difficulty ambulating, and decreased range of motion.No revision procedure has been scheduled.Attempts to obtain additional information have been made; however, no more is available at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2022-00956, 0001822565-2022-00957, and 0001822565-2022-00959.Gender solutions male (gsm) femoral component porous size 5, right catalog#: 00541201802 lot#: 63505961.Articular surface ultracongruent size 3 4 catalog#: 00542802309 lot#: 64063915.Metal backed patella component porous size 3 catalog#: 00541800003 lot#: 63943783.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.Medical records received identified severe three-compartment degenerative arthritis, noted erosion, osteophytes, hypertrophic synovium in all three compartments of the knee and a four millimeter resection from the femur due to knee flexion contracture preoperatively.No intraoperative complications were identified.The device history records were reviewed and no discrepancies were identified.A definitive root cause could not be determined with the information provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Search Alerts/Recalls
|