DAVOL INC., SUB. C.R. BARD, INC. -1213643 BARD SOFT MESH; SURGICAL MESH Back to Search Results
Catalog Number 0117014
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As reported, the bard soft mesh pre-shaped tore during implant.The subject device was not returned for evaluation, however a photo was provided.Evaluation of the photo finds that the surgeon tailored the mesh creating a v shape which shows cutting from the edges inward and then pulling of the mesh into the v shape.It also appears the doctor created a keyhole above the v shape of the mesh.Based on the photo evaluation and investigation performed, the mesh tearing cannot be confirmed as the mesh was tailored for use.Hence, no conclusion can be made.Review of manufacturing records confirms product was manufactured to specification, with no indication of a manufacturing related cause for the event reported.To date, this is the only reported complaint for this manufacturing lot of (b)(4) units released for distribution in november, 2020.The instructions-for-use (ifu) supplied with the mesh adequately describes the proper technique for implantation.The precautions section of the ifu states, "intact bard soft mesh exhibits high burst and tensile strength.However, when custom tailoring, in special circumstances where excessive force is placed on the mesh, the following guidelines may be helpful: when cutting a notch in the mesh, a v-shape with a radiused point will withstand more force than a v-shape which comes to a sharp point." not returned.
 
Event or Problem Description
As reported, during lichtenstein herniorrhaphy on (b)(6) 2022, the bard soft mesh was attempted to implant into the patient.As reported, the surgeon tried to pick the mesh up and the mesh broke with a hole just with a finger under it.It was also reported that there was no tear/hole on the mesh identified when opening the package or prior to insertion into the patient.As reported, the surgeon used another bard soft mesh to complete the procedure and there was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD SOFT MESH
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED -3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
8005566756
MDR Report Key14014452
Report Number1213643-2022-00134
Device Sequence Number6288525
Product Code FTL
UDI-Device Identifier00801741030918
UDI-Public(01)00801741030918
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K052155
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/28/2025
Device Catalogue Number0117014
Device Lot NumberHUEX1052
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/17/2022
Initial Report FDA Received Date04/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/30/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age58 YR
Patient SexMale
Patient Weight56 KG
-
-