• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL OG HELICAL XTRSOFT COIL 2X8; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL OG HELICAL XTRSOFT COIL 2X8; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Model Number 640HX0208
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Procode: krd/hcg.Initial reporter phone: (b)(6).The device is available to be returned for evaluation and testing.However, it has not been received to date as indicated as ¿other¿ in this section as the reason for non-evaluation.If the device returns, a device investigation will be performed.This is one of two products involved with the complaint and the associated manufacturer report numbers are 3008114965-2022-00256.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
 
Event or Problem Description
As reported by the field, during a coil embolization, a 2x6 orbit galaxy coil (640hx0206, 30574615) could not be detached, the physician raised pressure to red area, but the coil still could not detach.The second coil, a 2x8 galaxy xtrasoft coil (640hx0208, 30546086) had the same issue.The third device was used to complete the surgery.There was no patient injury reported.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Updated sections on this medwatch: b4, g3, g6, h2, h6 and h10.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.Complaint conclusion: as reported by the field, during a coil embolization, a 2x6 orbit galaxy coil (640hx0206, 30574615) could not be detached, the physician raised pressure to red area, but the coil still could not detach.The second coil, a 2x8 galaxy xtrasoft coil (640hx0208, 30546086) had the same issue.The third device was used to complete the surgery.There was no patient injury reported.The device was not returned for analysis; therefore, no further investigation can be performed.A manufacturing record evaluation was performed for the finished device 30546086 number, and no non-conformances related to the malfunction were identified.Failure to detach is a known potential issue associated with the use of this device.The instructions for use (ifu) contains several precautions related to this issue and includes instructions for troubleshooting the situation should it be encountered during use.Based on the minimal information provided, it is not possible to determine the root cause of the detachment failure.It was not reported whether the control cable or dcb were replaced during the procedure.It is also unknown if a pre-deployment electrical check was performed.The instructions for use (ifu) advises to verify that the microcoil delivery system is fully connected, and no faults are indicated on the dcb.If a fault exists, all connections between the dpu, the dcb, and the connecting cable should be reseated.If a fault still persists, the connecting cable should be replaced.If this does not correct the error, the dcb should be replaced.If the microcoil delivery system still continues to have a fault, the microcoil system should be retrieved and replaced with a new microcoil system.After depressing the detach button on the dcb or control cable (if the system is free of faults), the dcb should be replaced if the light and audible tone do not activate.Also, if a fault light is detected, the system ready light is not illuminated, or there is no detachment after two detachment attempts, the dcb should be replaced.If detachment still does not occur, the microcoil system should be carefully withdrawn and replaced.As part of the cerenovus quality process, all devices are manufactured, inspected, and released to approved specifications.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Updated sections on this medwatch: b4, d9, g3, g6, h2, h3, h6 and h10.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.Complaint conclusion updated with device analysis: as reported by the field, during a coil embolization, a 2x6 orbit galaxy coil (640hx0206, 30574615) could not be detached, the physician raised pressure to red area, but the coil still could not detach.The second coil, a 2x8 galaxy xtrasoft coil (640hx0208, 30546086) had the same issue.The third device was used to complete the surgery.There was no patient injury reported.A non-sterile og helical xtrsoft coil 2x8 was received contained in the decontamination pouch.Upon receiving the device, a visual inspection was performed, and no apparent damage was observed.Further inspection under microscopic magnification revealed that the embolic coil remains attached to the rest of the device.It can be noted inside the introducer.No other damage was found on the device.The functional test was performed, a lab sample prowler select plus was flushed with a lab sample syringe then the og helical xtrsoft coil 2x8 was introduced, and the device was purged into the blue zone; after that, the pressure gauge was increased to the green zone and before reaching this point the coil was detached without any difficulty.A manufacturing record evaluation was performed for the finished device 30546086 number, and no non-conformances related to the malfunction were identified.Failure to detach is a known potential issue associated with the use of this device.The instructions for use (ifu) contains several precautions related to this issue and includes instructions for troubleshooting the situation should it be encountered during use.The customer complaint regarding a coil failure to detach was not able to be confirmed based on the findings during the functional test, procedural factors may contribute to the reported failure since no abnormalities were found in the device that contribute to the complaint.A manufacturing record evaluation was performed, and no non-conformances related to the reported complaint condition were identified.It should be noted that product failure could be caused by multiple factors.It was not reported whether the control cable or dcb were replaced during the procedure.It is also unknown if a pre-deployment electrical check was performed.The instructions for use (ifu) advises to verify that the microcoil delivery system is fully connected, and no faults are indicated on the dcb.If a fault exists, all connections between the dpu, the dcb, and the connecting cable should be reseated.If a fault still persists, the connecting cable should be replaced.If this does not correct the error, the dcb should be replaced.If the microcoil delivery system still continues to have a fault, the microcoil system should be retrieved and replaced with a new microcoil system.After depressing the detach button on the dcb or control cable (if the system is free of faults), the dcb should be replaced if the light and audible tone do not activate.Also, if a fault light is detected, the system ready light is not illuminated, or there is no detachment after two detachment attempts, the dcb should be replaced.If detachment still does not occur, the microcoil system should be carefully withdrawn and replaced.As part of the cerenovus quality process, all devices are manufactured, inspected, and released to approved specifications.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OG HELICAL XTRSOFT COIL 2X8
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CODMAN & SHURTLEFF, INC. (JUAREZ)
circuito interior norte #1820
parque industrial salvarcar
juarez chihuahua 32574
MX   32574
Manufacturer Contact
michelleann garcia
31 technology dr
irvine, CA 92618
6465917981
MDR Report Key14014875
Report Number3008114965-2022-00257
Device Sequence Number4558810
Product Code KRD
UDI-Device Identifier10886704030737
UDI-Public10886704030737
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K093973
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2023
Device Model Number640HX0208
Device Catalogue Number640HX0208
Device Lot Number30546086
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/17/2022
Supplement Date Received by Manufacturer06/03/2022
08/01/2022
Initial Report FDA Received Date04/05/2022
Supplement Report FDA Received Date06/06/2022
08/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/18/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
OG HELICAL XTRSOFT COIL 2X8
Patient SexUnknown
-
-