ZIMMER ORTHOPAEDIC MFG. LTD. ART. SURFACE FIXED BEARING ULTRACONGRUENT (UC) LEFT 12 MM HEIGHT; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Unstable (1667)
Health Effect - Clinical Codes Pain (1994); Ambulation or Postural Difficulties (2544)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2022-00953, 0001822565-2022-00954, and 0001822565-2022-00955.Concomitant medical products: femur trabecular metal cruciate retaining (cr) standard porous catalog#: 4250280700,1 lot#: 64116357.Natural tibia tm two-peg porous fixed bearing left size h catalog#: 4253000830,1 lot#: 63927723.Metal backed patella component porous size 3 catalog#: 00541800003, lot#: 63976176.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported patient is experiencing pain, ambulation difficulties, and instability of the knee approximately 3 years post implantation.No revision procedure has been scheduled at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was unable to be confirmed due to limited information received from the customer.Initial medical records were provided and reviewed by a health care professional.Review of the available records identified no intra complication.No other records were provided.Device history record was reviewed and no discrepancies relevant to the reported event were found.Review of complaint history found no additional related issues for this item and the reported item and lot combination.The reported products were reviewed for compatibility and it was determined that the following implants are not compatible: patella and femur.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.A summary of the investigation has been sent to the complainant.A summary of the investigation was sent to the surgeon conveying proper surgical technique and use of the device.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ART. SURFACE FIXED BEARING ULTRACONGRUENT (UC) LEFT 12 MM HEIGHT
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14015124
Report Number3007963827-2022-00083
Device Sequence Number6274505
Product Code JWH
UDI-Device Identifier00889024640726
UDI-Public(01)00889024640726(17)250531(10)63659262
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K172524
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2025
Device Model NumberN/A
Device Catalogue Number42511200612
Device Lot Number63659262
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/08/2022
Supplement Date Received by Manufacturer05/17/2022
07/27/2022
Initial Report FDA Received Date04/05/2022
Supplement Report FDA Received Date06/10/2022
07/29/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/06/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
Patient Weight100 KG
Patient RaceWhite
-
-