DALE MEDICAL PRODUCTS, INC 270 STABILOCK ET HOLDER Back to Search Results
Model Number 270
Health Effect - Clinical Code Skin Discoloration (2074)
Date of Event 02/10/2022
Type of Reportable Event Serious Injury
Event or Problem Description
"patient has a small indentation on the upper lip from et tube holding device.Slight discoloration is seen on the outside of the upper lip.".
 
Additional Manufacturer Narrative
It was reported that a patient being treated using the dale 270 et tube holder has small indentation on the upper lip from et tube holding device.Slight discoloration is seen on the outside of the upper lip.Dale medical has investigated the report including inspection and testing of a device from the same lot number 27608 as the one reported with no issues noted.A complete review of the device history records for this production lot 27608 also found no issues.Dale medical also performed inspection and testing and dhr review on other production lot's that were produced in the same timeframe with no issues noted.Dale medical has reviewed our customer complaint files and noted that we had no other complaints against the dale 270 device in the last three years; 2019-2021, in that time frame dale has shipped (b)(4) dale 270 devices worldwide.The dale 270 device packaging for the adhesive base that is applied to the patient's skin contains this wording: apply skin prep.Let dry.To prevent skin ulcerations or skin tears, assess patient skin and medical conditions prior to applying.Monitor daily or more frequently.We have found through many years of experience with our 270 device that when all provisions of our instructions for use are followed, including using the supplied skin prep (barrier wipe), there are very few issues reported.Dale medical therefore has concluded that there appears to be no issues with the device and the cause of the small indentation on the upper lip and slight discoloration on the outside of the upper lip could be related to not using the supplied skin prep (barrier wipe), as the reporting facility had no record of the skin prep being used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
270 STABILOCK ET HOLDER
Common Device Name
270 STABILOCK ET HOLDER
Manufacturer (Section D)
DALE MEDICAL PRODUCTS, INC
40 kenwood circle
suite 7
franklin 02038 3298
Manufacturer (Section G)
DALE MEDICAL PRODUCTS, INC.
40 kenwood circle
suite 7
franklin 02038 3298
Manufacturer Contact
raymond gruneberg
40 kenwood circle
suite 7
franklin, MA 02038-3298
5083164608
MDR Report Key14015362
Report Number1214422-2022-00009
Device Sequence Number20317829
Product Code CBH
UDI-Device Identifier00817612020506
UDI-Public00817612020506
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
LISTING ONLY
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date12/17/2025
Device Model Number270
Device Catalogue Number270
Device Lot Number27608
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/11/2022
Initial Report FDA Received Date04/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/25/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexPrefer Not To Disclose
-
-