• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. WM CLARIS CPU; COMPUTER, DIAGNOSTIC, PROGRAMMABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. WM CLARIS CPU; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number H700977
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/23/2022
Type of Reportable Event Malfunction
Event or Problem Description
During device preparation for the paroxysmal supraventricular tachycardia procedure, while the patient was prepped, the workmate claris was powered on and there was no display which caused a delay.The conversion cable between the display and dvi was replaced, the dvi cable was replaced, the connections were checked, the reception settings were checked, the dws was restarted and discharged, but the issue persisted.The workmate claris was replaced and the procedure was completed with no adverse patient consequences.
 
Additional Manufacturer Narrative
One claris display workstation (dws) z440 was received for evaluation.Visual inspection revealed the front panel ports, chassis, and labels were free of physical damage.Power was applied to the dws, and the dws would not boot.The monitors would not function and no splash screen was observed.Over time the fans increased to maximum, but no other output signals happened.Based on replacement parts this symptom was isolated to the motherboard.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the investigation and information provided to abbott, the reported event was confirmed, and the root cause was attributed to the motherboard.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WM CLARIS CPU
Common Device Name
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key14015857
Report Number2184149-2022-00093
Device Sequence Number6300953
Product Code DQK
UDI-Device Identifier05415067009777
UDI-Public05415067009777
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K151911
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberH700977
Device Catalogue NumberH700977
Device Lot Number6494622
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/01/2022
Supplement Date Received by Manufacturer06/06/2022
Initial Report FDA Received Date04/05/2022
Supplement Report FDA Received Date06/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/21/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-