| Model Number |
70-10770 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Needle Stick/Puncture (2462)
|
| Date of Event |
03/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A supplier corrective action request (scar) was issued to the needle supplier becton - dickinson.The sample has not been returned to deroyal for evaluation at this time.An inventory check was made by deroyal, a total of 50 needles were inspected, and no discrepancies were identified during the inspection.This investigation is ongoing at this time.No further information is available at this time.We will provide follow-up report if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
Having breakage trouble with the item 5-9043 needle within this tray.No injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
A supplier corrective action request (scar) was issued to the needle supplier becton - dickinson.Representative samples were received and evaluated.Root cause: the supplier becton - dickinson was unable to determine the true root cause.Visual inspection was performed on the representative samples with a 10x magnifier lens.No defects or imperfections were observed.Then an inspection was performed with a 30x microscope.No defects or imperfections were observed.Corrective action: a corrective action has not been taken due to the root cause determination.A trend analysis was done for this product by becton-dickinson.This is the only complaint for this symptom from this and any other customer during the last 3 years.This investigation is complete at this time.No further information is available at this time.We will provide follow-up report if additional information becomes available.
|
| |
|
Search Alerts/Recalls
|