| Model Number |
406804 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During the procedure, a fluid leak was noted at the level of the non-return valve of the sheath resulting in a liquid entry into the stereotaxis systemr.The procedure continued with no consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
One 8.5f swartz braided introducer sheath was received for evaluation.Functional testing confirmed a fluid leak in the hemostasis valve.The cap was removed from the hemostasis hub and the hemostasis seals were microscopically inspected.Tearing, resulting in a hole, was noted in the proximal and distal seals.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the torn seals and subsequent leak remains unknown.The ifu states damage to the valve assembly may occur if inner catheter is withdrawn rapidly.
|
| |
|