• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION LOCKING TITANIUM ADAPTER; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION LOCKING TITANIUM ADAPTER; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE Back to Search Results
Catalog Number 5C4129
Medical Device Problem Codes Break (1069); Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/11/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Initial reporter facility name - (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that the packaging of a titanium adapter was ¿broken¿.This was observed before use of the device for peritoneal dialysis therapy.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.The returned photograph was reviewed, and the photograph showed the front of a pouch over labelled with a chinese text label indicating that the product was repacked and relabeled in china.It is not possible to determine if the package is broken or damaged based on the evaluation of the photograph.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
Correction made to a1: patient identifier: none (previously submitted as unknown).Correction made to b5: there was no patient involvement (previously submitted as there was no report of patient injury or medical intervention associated with this event) additional information was added to d9, h3, h6 and h10.The sample was received for evaluation.A visual inspection with the naked eye noted the pouch was over labelled with a chinese text label indicating that the product was repacked and relabeled in china.There was a slit (approximately 2mm) on the front of the pouch.The reported condition was verified.The cause of the condition could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING TITANIUM ADAPTER
Common Device Name
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - SWINFORD
foxford road
swinford
EI  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key14016050
Report Number1416980-2022-01589
Device Sequence Number4609940
Product Code KDJ
UDI-Device Identifier00085412007403
UDI-Public(01)00085412007403
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K152675
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/28/2025
Device Catalogue Number5C4129
Device Lot Number20K05H93
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/11/2022
Supplement Date Received by Manufacturer04/11/2022
07/05/2022
Initial Report FDA Received Date04/05/2022
Supplement Report FDA Received Date05/07/2022
07/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-