| Model Number |
RBY2C0410-B |
| Medical Device Problem Codes |
Material Deformation (2976); Premature Separation (4045)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Evaluation of the first returned ruby coil confirmed kinks on the pusher assembly.If the ruby coil is inadvertently mishandled at an extreme angle during advancement, damage such as this may occur.Further evaluation of the device revealed that the embolization coil was intact with the pusher assembly; therefore, the reported ruby coil detachment during the procedure could not be confirmed.Further evaluation also revealed offset coil winds.This damage was likely incidental to the complaint.Penumbra products are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2022-00153.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the left arteriovenous fistula (avf) using ruby coils and a non-penumbra microcatheter.During the procedure, while advancing a ruby coil into the target vessel using the microcatheter, the hospital technician bent the pusher assembly of the ruby coil and, consequently, the ruby coil unintentionally detached.Therefore, the physician used a syringe to aspirate and remove the detached ruby coil.The same issue then occurred with the next ruby coil.Therefore, the physician removed the microcatheter containing the detached ruby coil and the coil was flushed out.The procedure was completed using new ruby coils and the same microcatheter.There was no report of an adverse effect to the patient.
|
| |
|
Search Alerts/Recalls
|
|
|