| Model Number |
68LP-440 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "right side deflation." device remains implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "right side deflation." device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: opening in the posterior , crease fold, white particle inside the device and wear abrasion.Leak test was performed and found opening on posterior.A microscopic analysis was performed which identify crease smooth opening on posterior.The fill test inspection was performed, the result is no blockage.Based on the device analysis the final assessment is: a crease smooth opening on posterior assessed to a fold flaw opening.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "right side deflation." device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data.
|
| |
|
Search Alerts/Recalls
|