AMO PUERTO RICO MFG. INC. TECNIS IOL; EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS Back to Search Results
Model Number ZXT150
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/07/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for analysis therefore, a visual analysis of the complaint device cannot be completed.A review of the device/lot history record and historical data analysis for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch report will be filed.An attempt was made to obtain missing information; however, the account was not willing to provide patient information.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported during post-operative examination a scratch was found on the intraocular lens (iol).The doctor will be explanting the lens and replacing it with the same model lens.Through follow-up additional information was received confirming the iol was explanted (b)(6)2022.There was no unplanned incision enlargement, no serious patient injury, no unplanned suture(s), and no unplanned vitrectomy.Patient outcome post-lens exchange was reported as no issues.No further information is available.
 
Additional Manufacturer Narrative
Additional information: section d9: device available for evaluation: yes.Section d9: date returned to manufacturer: 27-apr-2022.Section h3: device evaluated by manufacturer: yes.Device evaluation: the complaint lens was received inside of a specimen cup.Visual inspection under magnification revealed that the lens was received cut in half with lens damage.The lens was cleaned and, no further issues, including cosmetic issues and scratches away from the cut portion of the lens were observed.The complaint issue "cosmetic issues"" was identified during product evaluation; however, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.Conclusion: based on the investigation results, there is no indication of a product malfunction or product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
EXTENDED DEPTH OF FOCUS INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14017105
Report Number3012236936-2022-00897
Device Sequence Number11901869
Product Code POE
UDI-Device Identifier05050474605206
UDI-Public(01)05050474605206(17)231016
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/16/2023
Device Model NumberZXT150
Device Catalogue NumberZXT150U200
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/10/2022
Supplement Date Received by Manufacturer05/20/2022
Initial Report FDA Received Date04/05/2022
Supplement Report FDA Received Date05/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/16/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-