HANGZHOU BEVER MEDICAL DEVICES CO., LTD. BD READY TO USE INTERMITTENT CATHETER Back to Search Results
Model Number RTU16M
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 03/09/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the whole week patient had an infection and wanted to be back to the original catheters.It was stated that patient was having trouble with coude tip and wanted straight tip catheters.It was unknown what medical intervention was reported.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that the whole week patient had an infection and wanted to be back to the original catheters.It was stated that patient was having trouble with coude tip and wanted straight tip catheters.It was unknown what medical intervention was reported.
 
Additional Manufacturer Narrative
The reported event was inconclusive.No sample was returned for evaluation.A potential failure mode for this is event would be "biocompatibility issues " and a potential root cause for this failure could be," unsuitable material".The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "contraindications the product is forbidden for use if the patient has acute urethritis, acute prostatitis, acute epididymitis, and/or acute urinary tract bleeding or injury.Warnings to help reduce the potential risk of infection and/or other complications, do not re-use.If discomfort or any sign of trauma occurs, discontinue use immediately and consult your doctor." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device was not returned.
 
Event or Problem Description
It was reported that the whole week patient had an infection and wanted to be back to the original catheters.It was stated that patient was having trouble with coude tip and wanted straight tip catheters.It was unknown what medical intervention was reported.
 
Additional Manufacturer Narrative
The reported event was inconclusive.No sample was returned for evaluation.A potential failure mode could be "biocompatibility issues".Therefore, a potential root cause for this failure could be "unsuitable material".The lot number was unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "contraindications : the product is forbidden for use if the patient has acute urethritis, acute prostatitis, acute epididymitis, and/or acute urinary tract bleeding or injury." "warnings: to help reduce the potential risk of infection and/or other complications, do not re-use.If discomfort or any sign of trauma occurs, discontinue use immediately and consult your doctor"."precautions: please consult your doctor before using this product if any of the following conditions are present: fever, chills, back pain, discomfort on emptying the bladder, severed urethra, unexplained urethral bleeding, pronounced stricture, false passage, urethritis - inflammation of the urethra, prostatitis - inflammation of the prostate gland, epididymitis - inflammation of the epididymis (testicle tube)." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD READY TO USE INTERMITTENT CATHETER
Common Device Name
INTERMITTENT CATHETER
Manufacturer (Section D)
HANGZHOU BEVER MEDICAL DEVICES CO., LTD.
no. 8-1, longquan road
cangqian town, yuhang district
hangzhou, zhejiang 31112 1
CH  311121
MDR Report Key14017145
Report Number1018233-2022-01927
Device Sequence Number17382842
Product Code GBM
UDI-Device Identifier06948796203974
UDI-Public(01)06948796203974
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K192468
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Importer
Report Source Other,Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/04/2022
Report Date (Section F) 03/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberRTU16M
Device Catalogue NumberRTU16M
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date03/09/2022
Event Location Other
Date Report to Manufacturer03/14/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer05/13/2022
07/01/2022
Initial Report FDA Received Date04/05/2022
Supplement Report FDA Received Date06/03/2022
07/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-