• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-3023-53
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/09/2022
Event Type  malfunction  
Event Description
It was reported that the cs300 intra-aortic balloon pump (iabp) was not holding the battery charge.It is unknown the circumstances under which the event occurred.It is also unknown if there was a patient involvement.However, there was no adverse event reported.
 
Manufacturer Narrative
A supplemental report will be submitted upon receipt of additional information or completion of our investigation.
 
Manufacturer Narrative
Testing of actual/suspected device|10) a getinge field service engineer (fse) evaluated the iabp unit and found that the batteries were past due for replacement.To fix the issue the fse replaced the batteries and then performed a full pm with calibration, functional and safety checks to meet factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that during a routine check it was observed that the cs300 intra-aortic balloon pump (iabp) was not holding the battery charge.It is unknown if there was a patient involvement.However, there was no adverse event reported.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CS300 INTRA-AORTIC BALLOON PUMP, ENGLISH, 110V
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key14017208
MDR Text Key288663422
Report Number2249723-2022-00739
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567107882
UDI-Public10607567107882
Combination Product (y/n)N
PMA/PMN Number
K063525
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 11/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-3023-53
Device Catalogue Number0998-00-3023-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received11/01/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/18/2016
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-