Model Number B2080-040
Medical Device Problem Codes Entrapment of Device (1212); Difficult to Remove (1528); Material Rupture (1546); Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/16/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the procedure was to treat a heavily calcified, mildly tortuous, de novo lesion in the right common femoral artery.After an unspecified stent was implanted in the right common femoral artery, post-dilatation was being performed with an 8x40mm armada 35 balloon dilatation catheter (bdc), when the balloon ruptured due to the heavy calcification at the lesion.The bdc was then being removed from the anatomy when it became stuck with the tip of the sheath.It was decided that the bdc and sheath would be removed from the lesion as a single unit, however the balloon became stuck in calcium, while the sheath was removed intact.The bdc became separated, and the tip of the device was left inside the vessel.The patient was put under general anesthesia (ga), and a cut down procedure was performed to successfully remove the ruptured balloon from the anatomy.There was no adverse patient sequela, however there was a reported clinically significant delay due to the cut down procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device is expected to be returned for evaluation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
 
Additional Manufacturer Narrative
Visual inspection was performed on the returned device.The reported balloon rupture and separation were confirmed.The reported difficulty removing the device could not be replicated in a testing environment due to the condition of the returned device.The reported entrapment of device could not be replicated in a testing environment as it was based on operational circumstances.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation determined the reported balloon rupture, difficulty removing the device, balloon separation, entrapment of device, delay to treatment/therapy, surgical intervention, removal of foreign body and hospitalization appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to design, manufacture or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 35 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key14017391
Report Number2024168-2022-03579
Device Sequence Number14592389
Product Code LIT
UDI-Device Identifier08717648155437
UDI-Public08717648155437
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K111899
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Model NumberB2080-040
Device Catalogue NumberB2080-040
Device Lot Number10819G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/16/2022
Supplement Date Received by Manufacturer05/24/2022
Initial Report FDA Received Date04/05/2022
Supplement Report FDA Received Date06/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/19/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
Patient SexMale
-
-