Model Number 367341
Medical Device Problem Code Material Discolored (1170)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported when using the bd vacutainer® push button blood collection set the tubing material was discolored.This event occurred 6 times.The following information was provided by the initial reporter and translated to english.The customer stated: "the pipeline is not normal transparent color, and it is yellowish to the naked eye.".
 
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.There were multiple medical device types reported to be involved.The information for each 510(k) number is as follows: common device name: intravascular administration set.Medical device type: fpa.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d9: device available for eval yes, d9: returned to manufacturer on: 01-apr-2022.Investigation summary: bd received 3 photos and 1 sample from the customer in support of this complaint.The photos and sample were evaluated and the indicated failure of discolored tubing was observed.Bd is able to confirm the customer¿s reported failure with the sample received.Based on a review of batch records and investigation results, no root cause from manufacturing was identified as a contributor.The device history record was reviewed with no issues being identified.There were no related quality notifications.All process and final inspections comply with specification requirements.H3 other text : see h10.
 
Event or Problem Description
It was reported when using the bd vacutainer® push button blood collection set the tubing material was discolored.This event occurred 6 times.The following information was provided by the initial reporter and translated to english.The customer stated: "the pipeline is not normal transparent color, and it is yellowish to the naked eye.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer (Section G)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14017432
Report Number1024879-2022-00194
Device Sequence Number17276006
Product Code JKA
UDI-Device Identifier50382903673419
UDI-Public50382903673419
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K030573
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date09/30/2022
Device Model Number367341
Device Catalogue Number367341
Device Lot Number0252250
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/21/2022
Supplement Date Received by Manufacturer07/08/2022
Initial Report FDA Received Date04/05/2022
Supplement Report FDA Received Date08/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/08/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-