SISTEMAS MEDICOS ALARIS, S.A. DE C.V. BD ALARIS SMARTSITE EXTENSION SET; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Model Number 20041E
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.Device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.
 
Event or Problem Description
It was reported that 3 bd alaris smartsite extension sets experienced leakage.The following information was provided by the initial reporter: contrast medication leaked after connecting with the open end for power injection.
 
Additional Manufacturer Narrative
The following fields were updated due to additional information: d.9.Device available for eval?: yes.D.9.Returned to manufacturer on: 29-apr-2022.H.6.Investigation: three 20041e samples were received without packaging for investigation; residual fluid was present in the returned samples.The customer reported the affected product to be from lot 20075513, however analysis of the affected lot number determined that it was not related to the affected product.Further information provided by the customer indicates that the leakage was from the smartsite of the component during injection of contrast.A visual inspection of the 20041e products revealed no defect or damage which could have contributed to a report of this nature.Pressure testing was conducted by connecting the female luer of the 20041e products to a 50ml bd plastipak syringe from bd stock whilst occluding the male luer; no leakage was observed.The lot number was not available and therefore it is not possible to perform a review of the production documentation for this particular product.In this instance, a definitive root cause could not be identified as testing of the returned samples did not identify any product defects that could have contributed to the customer¿s experience.
 
Event or Problem Description
It was reported that 3 bd alaris smartsite extension sets experienced leakage.The following information was provided by the initial reporter: contrast medication leaked after connecting with the open end for power injection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS SMARTSITE EXTENSION SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key14017439
Report Number9616066-2022-00360
Device Sequence Number4531759
Product Code FPA
UDI-Device Identifier27613203021665
UDI-Public27613203021665
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K960280
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number20041E
Device Catalogue Number20041E
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/17/2022
Supplement Date Received by Manufacturer04/29/2022
Initial Report FDA Received Date04/05/2022
Supplement Report FDA Received Date05/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-