AMO MANUFACTURING NETHERLANDS TECNIS IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number ZCT300
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the intraocular lens (iol) had an issue.Additional information received revealed that the leading haptic of the intraocular lens (iol) was caught between the piston and optic upon attempted insertion and the haptic got pulled off.There was contact with patient's operative eye.Another lens of different model (a 3-piece lens) was placed in the sulcus.No other intervention was required.The original lens was discarded.No other information was provided.
 
Additional Manufacturer Narrative
The device was not returned for analysis as the lens was discarded.Therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14017726
Report Number3012236936-2022-00742
Device Sequence Number11936668
Product Code MJP
UDI-Device Identifier05050474553149
UDI-Public(01)05050474553149(17)220706
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 04/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/06/2022
Device Model NumberZCT300
Device Catalogue NumberZCT300U110
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/14/2022
Initial Report FDA Received Date04/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/06/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE 1MTEC30
Patient SexUnknown
-
-