Model Number 380677-17
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/07/2022
Event Type  malfunction  
Manufacturer Narrative
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported complaint.The reported complaint was confirmed based on the field evaluation.The fse was able to recalibrate the master tool manipulator (mtm).The issue was resolved.The system was tested and verified as ready for use following the fse visit.A review of the site's system logs for the reported procedure date was conducted by rpms analyst.Investigation revealed there were no related system errors to have occurred during the surgical procedure that would have likely caused or contributed to the reported complaint.In addition, the system serial number and event date provided by the customer were confirmed.No image or video clip for the reported event was submitted for review.This complaint is considered a reportable event due to the following conclusion: the surgeon side console (ssc) was abandoned after the start of the procedure and the surgeon was able to continue with the procedure robotically by using the second ssc.While there was no harm or injury to the patient, the reported failure mode could likely cause or contribute to an adverse event if it were to recur.System unavailability after start of a surgical procedure could lead to the procedure to be converted/aborted and may lead to an injury due to the patient's inability to tolerate a conversion/abortion.
 
Event Description
It was reported that during a da vinci-assisted distal pancreatectomy and splenectomy surgical procedure, the right hand master tool manipulator (mtm) clutch button was not functioning as expected.Technical support was contacted and the technical service engineer (tse) was informed by the customer that if the mtm gimbal was flipped, the other clutch button did work, leaving one button inoperable.The surgeon moved to a secondary console in the room and proceeded with the case.The customer requested the field service engineer (fse) to follow up.The procedure was continuing as planned with no reported injury.Intuitive surgical, inc.(isi) attempted to obtain additional information related to the reported event, but no further details were available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAVINCI XI
Type of Device
SURGEON SIDE CONSOLE, SMART PEDALS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key14017944
MDR Text Key298332075
Report Number2955842-2022-10900
Device Sequence Number1
Product Code NAY
UDI-Device Identifier00886874110744
UDI-Public(01)00886874110744
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131861
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date 03/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number380677-17
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/07/2022
Initial Date FDA Received04/05/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/15/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
-
-