Model Number 83786
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device evaluated by mfr: the complaint device was received for product analysis.The main coil and the introducer sheath were returned for this complaint.The main coil was inspected, and it was found kinked and stretched.Blood was noticed in the main coil and inside the introducer sheath.Introducer sheath was inspected, and no anomalies were found.Twist lock was opened.No more damages were found.Microscopic inspection of the main coil was performed and revealed that the zap tip has a smooth surface.The interlocking arm was inspected and no anomalies were noted.Dimensional inspection of the main coil was performed and there were missing fiber bundles on the main coil due to coil stretching condition.The remaining measured dimensions were within specification.No other issues were identified during the product analysis.
 
Event or Problem Description
Reportable based on device analysis completed on 15mar2022.It was reported that the coil was difficult to insert.A 6mmx20cm interlock-35 coil was selected for use.During the procedure, it was noted that the device was difficult to insert in the catheter hub part.The coil was removed from the patient's body and the procedure was completed with a different device.There was no patient injury.However, device analysis revealed missing fiber bundles.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLOCK-35
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
jay johnson
two scimed place
maple grove, MN 55311
7634942574
MDR Report Key14017955
Report Number2134265-2022-03751
Device Sequence Number6301098
Product Code KRD
UDI-Device Identifier08714729793021
UDI-Public08714729793021
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K133208
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/14/2024
Device Model Number83786
Device Catalogue Number83786
Device Lot Number0028206879
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/15/2022
Initial Report FDA Received Date04/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CATHETER - MEDIKIT IMAGING CATHETER, 0.038INCH
Patient SexUnknown
-
-