| Model Number |
105-5095-000 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Codes |
Stroke/CVA (1770); Intracranial Hemorrhage (1891); Stenosis (2263); Foreign Body In Patient (2687)
|
| Date of Event |
03/24/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information that an apollo catheter ruptured with onyx and the patient had a stroke.The patient was undergoing onyx treatment for arteriovenous malformation (avm).The access vessel was the femoral artery. dsmo was injected. 23ml.Smooth injection.Pushed about.2ml of onyx and was seen coming out of proximal catheter.Everything pulled out.Patient then stroked.The catheter ruptured intact in the proximal section.The patient is alive with injury.Ancillary devices: unknown guide catheter, chikai guidewire.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received reported that apollo catheter rupture while placing the onyx caused non-targeted embolization and allowed for extravasation of onyx embolic material into the distal carotid or apparently the proximal middle cerebral artery causing significant narrowing and acute infarct.
|
| |
|
Event or Problem Description
|
|
Additional information was received indicating the lesion was in the right avm and the patient had moderate vessel tortuosity.No ac tions/interventions were taken to resolve the event.The device was prepared as indicated in the ifu.The procedure was not completed.There was no resistance upon injection of the liquid embolic agent into the apollo.The physician performed a slow push during in jection.The injection rate was followed as indicated in the ifu.There was no other damage to the catheter besides the rupture.The liquid embolic agent got embedded in an unintended location, in the left ica.It was unknown how the subject was recovering from the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: as found condition: the apollo micro catheter was returned for analysis inside a biohazard plastic bag and a shipping box.The onyx was not returned as it was consumed in the patient.Visual inspection/damage location details: no damages were found with the apollo distal marker/tip.Onyx was found solidified inside the micro catheter lumen and hub.The entire surface of the apollo micro catheter body was examined under magnification.The apollo micro catheter body was found to be ruptured at ~18.5 cm from the distal tip.Testing/analysis: the apollo usable length was measured to be ~165.0cm (specifications: 165.0cm ± 2.5cm).The catheter was flushed with water and found not patent as it was found to be occluded with the solidified onyx.Conclusion: based on the device analysis and reported information, the apollo micro catheter was confirmed to have "catheter ruptured" as the returned apollo micro catheter was found ruptured.The rupture appears to have occurred as a result of over-pressurization.Rupture can occur during injection of embolic material or when the distal portion of the catheter is kinked, prolapsed, or occluded.Rupture can also occur due to high injection rate, use of palm of hand pressure, increased injection force against resistance.There was no non-conformance to specifications identified that led to the reported issue.The review of lot history records shows that the finished device has met all manufacturing requirements and specifications during final assembly and quality inspection.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|