Model Number RT132 |
Device Problems
Connection Problem (2900); Infusion or Flow Problem (2964)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The complaint rt132 infant continuous flow breathing circuit is currently en route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow-up report upon completion of our investigation.
|
|
Event Description
|
A distributor in (b)(4) reported on behalf of a distributor in (b)(4) that the pressure line of an rt132 infant continuous breathing circuit would not connect to the ventilator.There was no patient involvment.
|
|
Manufacturer Narrative
|
(b)(4).Correction: h6 health effect - impact code incorrectly selected - changed to 2645.As the event occurred before patient use, there is no patient involvement.Method: the complaint rt132 infant continuous flow breathing circuit was returned to fisher & paykel healthcare (f&p) in new zealand for evaluation, where it was visually inspected and performance tested.Results: visual inspection of the rt132 infant continuous flow breathing circuit did not identify any damage or moulding defects.Performance testing of the returned circuit revealed the pressure line connected directly to the ventilator and generator as intended.Conclusion: we are unable to determine the cause of the reported event as no fault was found with the returned rt132 infant continuous flow breathing circuit.All rt132 infant continuous flow breathing circuits are visually inspected and pressure tested prior to being released for distribution, and those that fail are rejected.The subject rt132 infant continuous flow breathing circuit would have met the required specifications at the time of production.The user instructions that accompany the rt132 infant continuous flow breathing circuit states: "check all connections are tight before use." "perform a pressure and leak test on the breathing system and check for occlusions before connecting to a patient." ensure approptiate ventilator or flow source alarms are set before connecting breathing set to patient." "failure to comply with any of the following warnings may impair performance of the device or compromise safety (including potentially causing serious harm).".
|
|
Event Description
|
A distributor in italy reported on behalf of a distributor in france that the pressure line of an rt132 infant continuous flow breathing circuit would not connect to the ventilator.There was no patient involvement.
|
|
Search Alerts/Recalls
|