BARD PERIPHERAL VASCULAR, INC. ULTRASCORE 014; FOCUSED FORCE PTA BALLOON Back to Search Results
Model Number US1415064
Medical Device Problem Codes Difficult to Remove (1528); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 11/2023).
 
Event or Problem Description
It was reported that during an angioplasty procedure in the iliac artery via radial approach, the pta balloon was allegedly stuck upon removal from the patient.It was further reported that the balloon and the guidewire were removed as one unit.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one ultrascore 014 pta dilatation catheter was received for evaluation.During visual evaluation, no kinks were noted to the catheter shaft and no anomalies to the luers/y-body.During functional testing, the sample guidewire lumen was flushed successfully.The patency of the guidewire lumen was tested using an in-house guidewire and was unable to be inserted.The catheter was cut and blood was noted within the guidewire lumen.No other functional testing was performed.Therefore, the investigation is inconclusive for the reported difficult to remove as functional testing could not be performed for the sheath related issue.A definitive root cause for the alleged difficult to remove could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 11/2023).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during an angioplasty procedure in the iliac artery via radial approach, the pta balloon was allegedly stuck upon removal from the patient.It was further reported that the balloon and the guidewire were removed as one unit.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one ultrascore 014 pta dilatation catheter was received for evaluation.During visual evaluation, no kinks were noted to the catheter shaft and no anomalies to the luers/y-body.During functional testing, the sample guidewire lumen was flushed successfully.The patency of the guidewire lumen was tested using an in-house guidewire and was unable to be inserted.The catheter was cut and dried blood was noted within the guidewire lumen.No other functional testing was performed.Therefore, the investigation is inconclusive for the reported difficult to remove as functional testing could not be performed for the sheath related issue.The investigation is inconclusive for the reported difficult to remove and device-device incompatibility as functional testing could not be performed for the guidewire related issue due to dried blood within the guidewire lumen.A definitive root cause for the alleged difficult to remove and device-device incompatibility could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 11/2023), g3, h6 (device, method).H11: g1.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during an angioplasty procedure in the iliac artery via radial approach, the pta balloon was allegedly stuck upon removal from the patient.It was further reported that the balloon and the guidewire were removed as one unit.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASCORE 014
Common Device Name
FOCUSED FORCE PTA BALLOON
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
C.R. BARD, INC. (GFO)
289 bay road
queensbury 12804
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14019202
Report Number2020394-2022-00264
Device Sequence Number14966541
Product Code PNO
UDI-Device Identifier00801741114021
UDI-Public(01)00801741114021
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K172571
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberUS1415064
Device Catalogue NumberUS1415064
Device Lot NumberGFFX1479
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/14/2022
Supplement Date Received by Manufacturer05/05/2022
07/27/2022
Initial Report FDA Received Date04/06/2022
Supplement Report FDA Received Date05/09/2022
08/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/04/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
-
-