| Model Number |
RONYX30015UX |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
03/28/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An attempt was made to use one resolute onyx coronary drug eluting stent and one telescope guide extension catheter to treat a non-tortuous, moderately calcified lesion with 95% stenosis in the proximal of the circumflex (cx) artery.Both devices were inspected with no issues noted.The telescope was prepped per ifu with no issues noted.The stent was negatively prepped with no issues noted.The lesion was pre-dilated with a 3.0x12mm balloon.The devices did not pass through a previously-deployed stent.Resistance was encountered when advancing the stent.Excessive force was not used during delivery.Resistance was not encountered when advancing the telescope and excessive force was not used.It was reported that the stent failed to cross the lesion.The telescope was then used for extra support but the stent was still unsuccessful.Stent dislodgement occurred when removing the stent back into the telescope tip.The physician initially tried to use the stent balloon to remove the dislodged stent but this was unsuccessful.A 1.5mm balloon was then dilated inside the dislodged stent and the dislodged stent was successfully pulled back into the non-medtronic guide catheter and removed.The patient is alive with no further injury.It was stated that there was no alleged issue against the telescope device.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: there were no issues noted during insertion of the resolute onyx into the telescope gec.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|