Model Number RONYX30015UX
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Insufficient Information (4580)
Event Date 03/28/2022
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
An attempt was made to use one resolute onyx coronary drug eluting stent and one telescope guide extension catheter to treat a non-tortuous, moderately calcified lesion with 95% stenosis in the proximal of the circumflex (cx) artery.Both devices were inspected with no issues noted.The telescope was prepped per ifu with no issues noted.The stent was negatively prepped with no issues noted.The lesion was pre-dilated with a 3.0x12mm balloon.The devices did not pass through a previously-deployed stent.Resistance was encountered when advancing the stent.Excessive force was not used during delivery.Resistance was not encountered when advancing the telescope and excessive force was not used.It was reported that the stent failed to cross the lesion.The telescope was then used for extra support but the stent was still unsuccessful.Stent dislodgement occurred when removing the stent back into the telescope tip.The physician initially tried to use the stent balloon to remove the dislodged stent but this was unsuccessful.A 1.5mm balloon was then dilated inside the dislodged stent and the dislodged stent was successfully pulled back into the non-medtronic guide catheter and removed.The patient is alive with no further injury.It was stated that there was no alleged issue against the telescope device.
 
Manufacturer Narrative
Additional information: there were no issues noted during insertion of the resolute onyx into the telescope gec.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key14019441
MDR Text Key288653618
Report Number9612164-2022-01332
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier00643169556980
UDI-Public00643169556980
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 06/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/06/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberRONYX30015UX
Device Catalogue NumberRONYX30015UX
Device Lot Number0010657181
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/22/2022
Date Device Manufactured04/26/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age81 YR
Patient SexMale
-
-