• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE RX; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ULTRAVERSE RX; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number U415034RX
Medical Device Problem Codes Difficult to Remove (1528); Detachment of Device or Device Component (2907); Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an angioplasty procedure in the cephalic vein, the pta balloon allegedly got stuck in the guidewire and was difficult to remove when the physician adjusted the position of the balloon.It was further reported that the balloon and guidewire were removed together as one unit.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one ultraverse rx pta catheter was returned for evaluation.On visual evaluation, the balloon was noted to be detached from the catheter exposing the inner guide wire lumen and the detached balloon was not returned for evaluation.No other anomalies were noted.No functional testing was performed due to the condition of the device.Therefore, the investigation remains inconclusive for the reported difficult to remove and device-device incompatibility as no functional testing was performed due to the condition of the device returned for evaluation.The investigation is confirmed for the identified balloon detachment as the balloon was detached from the catheter exposing the inner guide wire lumen and the detached balloon was not returned for evaluation.A definitive root cause for the reported difficult to remove , device-device incompatibility and identified balloon detachment could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.(expiry date: 02/2024).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRAVERSE RX
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford N A
EI   N A
Manufacturer Contact
judy ludwig
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14020646
Report Number2020394-2022-00268
Device Sequence Number2631851
Product Code LIT
UDI-Device Identifier00801741056802
UDI-Public(01)00801741056802
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K131199
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberU415034RX
Device Catalogue NumberU415034RX
Device Lot NumberCMFQ0267
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/29/2022
Initial Report FDA Received Date04/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-