INTEGRA LIFESCIENCES SWITZERLAND SAR DIRECTLINK ICP MODULE; DIRECTLINK - MODULE Back to Search Results
Model Number 826828
Medical Device Problem Code Low Readings (2460)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
2 of 2 reports.Other mfg report number: 3013886523-2022-00135.A facility reported a directlink icp module was showing negative readings after implantation.The sensor was replaced and one hour surgical delay was reported because when the sensor was replaced, the negative readings persisted; therefore, the sensors were not the cause.The monitoring continued with a competitor module available.
 
Additional Manufacturer Narrative
The directlink icp module was returned for evaluation: dhr - conformed to the specifications when released to stock.Failure analysis - directlink module 823828 was tested according to procedure applicable, and no product problem was identified.Root cause - the root cause of the issue reported by customer could not be determined as the device worked correctly.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIRECTLINK ICP MODULE
Common Device Name
DIRECTLINK - MODULE
Manufacturer (Section D)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
rue girardet 29
le locle
SZ 
Manufacturer (Section G)
INTEGRA LIFESCIENCES SWITZERLAND SAR
rue girardet 29
ch-2400
le locle
SZ  
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key14021219
Report Number3013886523-2022-00136
Device Sequence Number19481883
Product Code GWM
UDI-Device Identifier10381780515784
UDI-Public(01)10381780515784
Combination Product (Y/N)N
PMA/510(K) Number
K152670
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number826828
Device Catalogue Number826828
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/21/2022
Supplement Date Received by Manufacturer06/21/2022
Initial Report FDA Received Date04/06/2022
Supplement Report FDA Received Date07/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/02/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexMale
-
-