| Model Number |
50-7210 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/02/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) initial emdr submission.A follow up emdr will be submitted if additional information becomes available.(b)(4).
|
| |
|
Event or Problem Description
|
|
When the nurse wanted to drain the patient with a bottle, the tubing (drainage line) fell off the bottle neck.She closed the clamp on the drainage line, took a new bottle and the same thing happened again.Was the drainage line disconnected from the bottle at any point during the drain or was it noticed prior to drainage? yes, during the drain and it was noticed.If it's before use and did they attempt to reconnect it? during use, and no attempt to reconnect.Need to verify the line was disconnected prior to use and the bottle was not used.Bottle was used.Was the clamp properly closed until after the plunger was pressed, all of our patients are trained.So this should be done correctly.Where is the catheter located? abdomen or chest? unknown.Is there a photo or sample available showing the reported issue? there is no phot and no sample available.What was the impact to the patient? no impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Pr (b)(4) follow-up emdr for device evaluation: no photos or samples were received by our quality team for evaluation therefore the failure mode could not be verified.A review of the internal manufacturing device records and raw material history files for the lot 0001358259 was performed and no recorded quality problems or rejections related to this incident were found.Product undergoes inspections throughout manufacturing, no issues related to the reported incident were identified, all procedural and functional requirements for product release have been met.Assembly of the drainage line onto the spike is performed manually.Once the adhesive is applied onto the drainage line, it is then connected onto the spike and personnel must hold the junction before the product moves onto the next stage of assembly.During our investigation it was identified that personnel were not following proper procedure after applying the adhesive onto the drainage line and were moving the product without holding the connection for the appropriate amount of time.Additionally, we identified that preventative maintenance was not properly performed on one of the machines which adheres the drainage line to the spike.Based on the quality team's investigation, it was determined that the disconnected drainage line was likely due to both personnel not following procedure and manufacturing equipment.Manufacturing personnel have been notified of this incident and additional training was provided.Updates will be implemented to ensure the proper maintenance is performed on the manufacturing equipment.Complaints received for this device and defect will continue to be monitored for future occurrences.H3 other text : see manufacturer here.
|
| |
|
Event or Problem Description
|
|
When the nurse wanted to drain the patient with a bottle, the tubing (drainage line) fell off the bottle neck.She closed the clamp on the drainage line, took a new bottle and the same thing happened again.Was the drainage line disconnected from the bottle at any point during the drain or was it noticed prior to drainage? yes, during the drain and it was noticed.If it's before use and did they attempt to reconnect it? during use, and no attempt to reconnect.Need to verify the line was disconnected prior to use and the bottle was not used.Bottle was used.Was the clamp properly closed until after the plunger was pressed.All of our patients are trained.So this should be done correctly.Where is the catheter located? abdomen or chest.Unknown.Is there a photo or sample available showing the reported issue? there is no phot and no sample available.What was the impact to the patient? no impact.
|
| |
|
Event or Problem Description
|
|
When the nurse wanted to drain the patient with a bottle, the tubing (drainage line) fell off the bottle neck.She closed the clamp on the drainage line, took a new bottle and the same thing happened again.Was the drainage line disconnected from the bottle at any point during the drain or was it noticed prior to drainage? yes, during the drain and it was noticed.If it's before use and did they attempt to reconnect it? during use, and no attempt to reconnect.Need to verify the line was disconnected prior to use and the bottle was not used.Bottle was used.Was the clamp properly closed until after the plunger was pressed all of our patients are trained.So this should be done correctly.Where is the catheter located? abdomen or chest unknown.Is there a photo or sample available showing the reported issue? there is no phot and no sample available.What was the impact to the patient? no impact.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) follow-up emdr for device evaluation: six physical samples were provided to our quality team for investigation.One of the bottles was returned in an opened package.The remaining five bottles were returned in sealed packages.Upon visual inspection of the opened package sample, it was observed that the drainage line was disconnected from the bottle therefore, the reported failure mode was confirmed.The remaining bottles were also inspected.No anomalies were observed.The drainage line is attached to the bottle.A review of the internal manufacturing device records and raw material history files for the lot 0001358259 was performed and no recorded quality problems or rejections related to this incident were found.Product undergoes inspections throughout manufacturing, no issues related to the reported incident were identified, all procedural and functional requirements for product release have been met.Assembly of the drainage line onto the spike is performed manually.Once the adhesive is applied onto the drainage line, it is then connected onto the spike and personnel must hold the junction before the product moves onto the next stage of assembly.During our investigation it was identified that personnel were not following proper procedure after applying the adhesive onto the drainage line and were moving the product without holding the connection for the appropriate amount of time.Additionally, we identified that preventative maintenance was not properly performed on one of the machines which adheres the drainage line to the spike.Based on the quality team's investigation, it was determined that the disconnected drainage line was likely due to both personnel not following procedure and manufacturing equipment.Manufacturing personnel have been notified of this incident and additional training was provided.Updates will be implemented to ensure the proper maintenance is performed on the manufacturing equipment.Complaints received for this device and defect will continue to be monitored for future occurrences.
|
| |
|