• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: K2M, INC. SCREW DRIVER W/ THREAD; SIZE 10; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

K2M, INC. SCREW DRIVER W/ THREAD; SIZE 10; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 8801-90034
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
A company representative reported that the tip of an ozark screwdriver fractured.The reported event was noted during inspection.There is no procedure associated with this reported event, no patient involvement, and no adverse consequence has been reported.
 
Additional Manufacturer Narrative
A visual inspection confirmed the reported issue of tip fracture.The threaded inner shaft tip was observed to be fractured off.Device history records were reviewed for the corresponding lot and no relevant manufacturing issues were identified.A review of complaint history associated with the subject lot and catalog number was performed and no adverse trends were observed.A definitive root cause could not be determined.Potential root causes include: not using torque limiting handle, excessive force, and hard bone density.
 
Event or Problem Description
A company representative reported that the tip of an ozark screwdriver fractured.The reported event was noted during inspection.There is no procedure associated with this reported event, no patient involvement, and no adverse consequence has been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCREW DRIVER W/ THREAD; SIZE 10
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer (Section G)
K2M, INC.
600 hope parkway se
leesburg VA 20175
Manufacturer Contact
rita karan
2 pearl court
allendale, NJ 07401
2017608000
MDR Report Key14021741
Report Number3004774118-2022-00130
Device Sequence Number6136440
Product Code LXH
UDI-Device Identifier10888857400863
UDI-Public10888857400863
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8801-90034
Device Catalogue Number8801-90034
Device Lot NumberJKKY
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/08/2022
Supplement Date Received by Manufacturer06/06/2022
Initial Report FDA Received Date04/06/2022
Supplement Report FDA Received Date07/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/11/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-