• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT GMBH ALINITY S CHAGAS; TRYPANOSOMA CRUZI (E COLI, RECOMBINANT) ANTIGEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GMBH ALINITY S CHAGAS; TRYPANOSOMA CRUZI (E COLI, RECOMBINANT) ANTIGEN Back to Search Results
Catalog Number 06P08-50
Medical Device Problem Code False Positive Result (1227)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer stated that in the month of (b)(6) 2022, there were 4 reactive alinity s chagas patient samples that were negative using the esa chagas method.When the customer encounters a reactive result on the alinity s processing module, they automatically repeat the sample again in duplicate and if it remains reactive, then they send it out for confirmation using the esa chagas method.The customer indicated that of 13 repeatedly reactive alinity s chagas samples that were sent out for confirmation, 4 of those were reported as confirmation negative (esa method).No adverse impact to donor or patient management was reported.
 
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.All available patient/donor information was included.Additional patient/donor details are not available.
 
Additional Manufacturer Narrative
Trending review determined no adverse trend for the issue for the product.Historical complaint review determined there is normal complaint activity for the lot number.Device history record review did not show any potential non-conformances, or deviations.Labeling was reviewed and found to adequately address the issue under review.A customer data review for alinity s chagas reagent, lot 26408be00 was performed.The analysis included initial reactive rates (irr), repeat reactive rates (rrr) and specificity (assuming zero prevalence) at rhode island blood center (ribc), across whole blood (wb) peer sites and all us customer sites using this lot.Aggregated performance of lot 26408be00 across all us customer sites (regardless of peer group) and selected wb peer sites is within product requirements and comparable to the performance of other lots of the same assay evaluated in the comparison.Additionally, a review of the overall monthly reactive rates from october 2021 to march 2022 was performed.The review demonstrates that reactive rates in january 2022 at ribc is within the range of historical performance.The performance of the assay across all months is within product requirements.Based on the investigation, no deficiency for lot number 26408be00 was identified.
 
Additional Manufacturer Narrative
Upon review on september 12, 2022, it was discovered that additional associated devices were omitted on the previous submission and have been added to this submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALINITY S CHAGAS
Common Device Name
TRYPANOSOMA CRUZI (E COLI, RECOMBINANT) ANTIGEN
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
Manufacturer (Section G)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM   65205
Manufacturer Contact
siobhan wright
lisnamuck
post market surveillance
longford N39 E-932
EI   N39 E932
433331157
MDR Report Key14021806
Report Number3002809144-2022-00119
Device Sequence Number11421328
Product Code QHM
UDI-Device Identifier00380740117221
UDI-Public00380740117221
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/16/2022
Device Catalogue Number06P08-50
Device Lot Number26408BE00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/17/2022
Supplement Date Received by Manufacturer06/09/2022
09/12/2022
Initial Report FDA Received Date04/06/2022
Supplement Report FDA Received Date06/10/2022
09/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/09/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ALNTY S SYSTEM, 06P16-01, AS1215; ALNTY S SYSTEM, 06P16-01, AS1215; ALNTY S SYSTEM, 06P16-01, AS1215; ALNTY S SYSTEM, 06P16-01, AS1233; ALNTY S SYSTEM, 06P16-01, AS1234
Patient SexUnknown
-
-