| Model Number |
M3702 |
| Medical Device Problem Code |
Device Contaminated During Manufacture or Shipping (2969)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown cardiovascular procedure on 12/27/2021 and suture was used.After the package was opened, there was a foreign matter stuck in the needle park foam.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Investigation summary: one opened straight-pack folder with four undispensed needle-sutures product code m3702 was returned for analysis.The product code should be multistrand and each straight-pack folder must contain four strands single-armed.As per visual analysis of the sample received a foreign matter (hair) caught in a section of the foam park was observed.Based on the information currently available, the foreign matter was identified during the investigation of the sample received.A manufacturing record evaluation was performed for the finished device lot, and no non-conformance's were identified.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|