U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL ADVANCED LOCKING SCREW IMN SCREWS Ø5X70MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL ADVANCED LOCKING SCREW IMN SCREWS Ø5X70MM; SCREW, FIXATION, BONE Back to Search Results
Model Number 2361-5070S
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Failure of Implant (1924); Perforation (2001); Insufficient Information (4580)
Date of Event 03/14/2022
Type of Reportable Event Serious Injury
Event or Problem Description
As reported: "on (b)(6) 2022, revision surgery was performed due to a t2 alpha advanced locking screw backing out of a tibial nail.".
 
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided on a supplemental report.Device retained by the surgeon.
 
Event or Problem Description
As reported: "on (b)(6) 2022, revision surgery was performed due to a t2 alpha advanced locking screw backing out of a tibial nail".
 
Additional Manufacturer Narrative
Please note correction to section b1.The reported event could be confirmed, since the migration of the screw is visible on one of the received pictures.The received pictures were reviewed.One picture with the complained screw was provided.Just the upper part of the screw is visible, there are no damages at the visible section of the screw, the reported lot number can be confirmed.One x-ray was received of the proximal end of the nail.Based on this x-ray the migration of the screw cannot be confirmed and no root cause can be defined because just one plane is visible and as no x-ray of the initial situation after the implantation was provided.One picture of the leg below the knee was provided, it is visible that the screw head of a locking screw has pierced the skin.Based on this picture the complained migration of the screw can be confirmed.The screw was not returned for evaluation, therefore no further evaluation is possible.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED LOCKING SCREW IMN SCREWS Ø5X70MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14022144
Report Number0009610622-2022-00116
Device Sequence Number2571479
Product Code HWC
UDI-Device Identifier07613327381504
UDI-Public07613327381504
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K193308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number2361-5070S
Device Catalogue Number23615070S
Device Lot NumberK02BCDE
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/14/2022
Supplement Date Received by Manufacturer05/24/2022
Initial Report FDA Received Date04/06/2022
Supplement Report FDA Received Date06/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/20/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age85 YR
Patient SexFemale
-
-