|
The reported event was inconclusive.No sample was returned for evaluation.A potential root cause for this failure could be due to "high modulus silicone or contact with a sharp object (i.E.,needle sticks, etc.) or exposure to petrolatum based lubricant or other degrading materials or pinhole in balloon or cuff".The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿warning: do not use ointments or lubricants having a petrolatum base.They will damage the catheter.Visually inspect the product for any imperfections or surface deterioration prior to use.¿ use luer tip syringe to inflate with stated ml of sterile water.Or ¿ for pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.For urological use only.Bard, bardia, bardex, uriplan, ez-lok, and lubricath are trademarks and/or registered trademarks of c.R.Bard, inc.Manufacturer: to deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.A new connector with a needle free sample port has been added to this product.Instructions for use for the needle-free sampling 1.Kink the drainage tubing at a minimum of 5cm below the sampling port.2.Wipe the surface of the port with an alcohol swab.3.Using an aseptic technique, position the syringe (luer slip tip only) in the centre, perpendicular to the surface of the port, and then press the tip of the syringe into the sampling port.4.Aspirate the desired volume and then remove the syringe.5.Wipe the surface of the port with an alcohol swab.6.Unkink the tubing and send the correctly labelled specimen to the laboratory.This is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable laws and regulations.¿ correction: g.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|