| Model Number |
GLM915040 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Aneurysm (1708)
|
| Date of Event |
08/05/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Information regarding patient height, race, and ethnicity was not reported.Patient identifier: (b)(6).Procode: krd/hcg.Device manufacture date: not available at time of reporting.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.This is one of four products involved with the complaint and the associated manufacturer report numbers are 3008114965-2022-00252, 3008114965-2022-00253, and 3008114965-2022-00254.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
|
| |
|
Event or Problem Description
|
|
As reported via (b)(6), a (b)(6)female (subject (b)(6)) with a past medical history of controlled hypertension underwent balloon-assisted coil embolization of a ruptured anterior communicating artery aneurysm with associated subarachnoid hemorrhage (sah) on (b)(6)2021.The patient experienced aneurysm recanalization on (b)(6)2022.Digital subtraction angiography (dsa) performed at the 180-day (6-month) follow-up visit on (b)(6)2022 showed modified raymond-roy classification score of class iiib: residual aneurysm with contrast along aneurysm wall.She subsequently underwent aneurysm retreatment due to residual aneurysm on (b)(6)2022.Immediate pre-procedure angiography on (b)(6)2021 revealed a ruptured anterior communicating artery aneurysm with the following dimensions: height 6mm, dome 4mm, maximum aneurysm diameter 6mm, neck size 4mm, and dome-to-neck ratio 1.0mm.Balloon-assisted coil embolization was subsequently performed on (b)(6)2021.A total of four cerenovus coils were implanted during the procedure: galaxy g3 7mm x 21cm (gly120721, 30470358), galaxy g3 xsft hel 2mm x 8cm (glx120208, 3045189), galaxy g3 xsft 4mm x 10cm (glx120410, 30504557) and galaxy g3 mini 1.5mm x 4cm (glm915040, 30490128) with 34% angiosuite packing density.The coils were delivered via an excelsior 10 (stryker) microcatheter.Heparin was administered during the procedure.The study coils were successfully implanted at the target site with final modified raymond-roy score of class i: complete = complete obliteration.In the opinion of the investigator, the treatment of the target aneurysm was considered complete.There were no reported intraoperative complications or study device deficiencies.There was no microcatheter kickback (loss of access to aneurysm) experienced.The parent vessel diameter was 2.3mm.The patient was discharged home with self-care on (b)(6)2021 with modified rankin scale (mrs) score of 0.A neuroform atlas stent (stryker) and six target coils (stryker) were implanted during the aneurysm recanalization.There were no intraoperative complications.The patient was discharged home with self-care on (b)(6)2022.The aneurysm recanalization was considered resolved on (b)(6)2022.Per the pi, the event was moderate in severity and possibly related to the study device.The event was considered serious as it required hospitalization.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Updated sections on this medwatch: b4, g3, g6, h2, h6 and h10.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.Complaint conclusion: as reported via the sterling study, a 67-year-old female (subject (b)(6)) with a past medical history of controlled hypertension underwent balloon-assisted coil embolization of a ruptured anterior communicating artery aneurysm with associated subarachnoid hemorrhage (sah) on (b)(6) 2021.The patient experienced aneurysm recanalization on (b)(6) 2022.Digital subtraction angiography (dsa) performed at the 180-day (6-month) follow-up visit on (b)(6) 2022 showed modified raymond-roy classification score of class iiib: residual aneurysm with contrast along aneurysm wall.She subsequently underwent aneurysm retreatment due to residual aneurysm on (b)(6) 2022.Immediate pre-procedure angiography on (b)(6) 2021 revealed a ruptured anterior communicating artery aneurysm with the following dimensions: height 6mm, dome 4mm, maximum aneurysm diameter 6mm, neck size 4mm, and dome-to-neck ratio 1.0mm.Balloon-assisted coil embolization was subsequently performed on (b)(6) 2021.A total of four cerenovus coils were implanted during the procedure: galaxy g3 7mm x 21cm (gly120721, 30470358), galaxy g3 xsft hel 2mm x 8cm (glx120208, 3045189), galaxy g3 xsft 4mm x 10cm (glx120410, 30504557) and galaxy g3 mini 1.5mm x 4cm (glm915040, 30490128) with 34% angiosuite packing density.The coils were delivered via an excelsior 10 (stryker) microcatheter.Heparin was administered during the procedure.The study coils were successfully implanted at the target site with final modified raymond-roy score of class i: complete = complete obliteration.In the opinion of the investigator, the treatment of the target aneurysm was considered complete.There were no reported intraoperative complications or study device deficiencies.There was no microcatheter kickback (loss of access to aneurysm) experienced.The parent vessel diameter was 2.3mm.The patient was discharged home with self-care on (b)(6) 2021 with modified rankin scale (mrs) score of 0.A neuroform atlas stent (stryker) and six target coils (stryker) were implanted during the aneurysm recanalization.There were no intraoperative complications.The patient was discharged home with self-care on (b)(6) 2022.The aneurysm recanalization was considered resolved on (b)(6) 2022.Per the pi, the event was moderate in severity and possibly related to the study device.The event was considered serious as it required hospitalization.The device remains implanted; therefore, no further investigation can be performed.A manufacturing record evaluation was performed for the finished device 30490128 number, and no non-conformances related to the malfunction were identified.Aneurysm recanalization is a condition requiring additional intervention or retreatment or can be reasonably expected to result in medical or surgical intervention, including hospitalization.In this case, intervention and hospitalization was indicated to prevent further injury or complications.Furthermore, the relationship of the study coils to the reported aneurysm recanalization cannot be excluded.As part of the cerenovus quality process, all devices are manufactured, inspected, and released to approved specifications.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.
|
| |
|
Search Alerts/Recalls
|
|