DATASCOPE CORP. - FAIRFIELD LINEAR 7.5 FR. 34CC IAB; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0684-00-0479-01C
Medical Device Problem Codes Leak/Splash (1354); Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/13/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that after approximately two days of intra-aortic balloon (iab) therapy the console generated a check iab catheter alarm.The customer checked the iab but found no issue.Two hours later, the console generated the same alarm and blood was seen in the tubing.The iab had to be removed by surgery in the intensive care unit (icu).There was no patient harm or adverse event reported.
 
Additional Manufacturer Narrative
(b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
 
Additional Manufacturer Narrative
The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.Complaint record id # (b)(4).
 
Event or Problem Description
N/a.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
The product was returned with the membrane completely unfolded and blood on the interior and exterior of the catheter and between the catheter and the sheath.A kink was found on the catheter tubing approximately 59.7cm from the iab tip.Additionally a second kink was found on the catheter tubing and inner lumen near the y-fitting approximately 76.5cm from the iab tip.The inner lumen was found occluded with dried blood.The occlusion was unable to be cleared.An underwater leak test of the balloon, catheter, y-fitting and extracorporeal tubing was performed and six leaks were detected on the membrane.One abrasion leak at approximately 1.8cm from the rear seal and measuring 0.013cm in length.Five sharp edge leaks ranging from 2.5cm to 7.4cm from the rear seal and measuring ranging from 0.127cm to 0.635cm in length.The reported problem was most likely triggered by a leak which was found on the membrane.Under magnification, a whitish patch was observed around the leak.This whitish patch is the typical appearance of an abrasion mark which is caused by calcified plaque in the aorta.The other five penetrations found appear to have been caused by a sharp object that may have occurred during iab removal.A lot history record review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Reference complaint #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR 7.5 FR. 34CC IAB
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer (Section G)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
Manufacturer Contact
brian schaeffer
15 law drive
fairfield, NJ 
MDR Report Key14022386
Report Number2248146-2022-00280
Device Sequence Number2632971
Product Code DSP
UDI-Device Identifier10607567109329
UDI-Public10607567109329
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K041281
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/14/2024
Device Model Number0684-00-0479-01C
Device Catalogue Number0684-00-0474
Device Lot Number3000179632
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/16/2022
Supplement Date Received by Manufacturer04/01/2022
07/22/2022
Initial Report FDA Received Date04/06/2022
Supplement Report FDA Received Date04/18/2022
07/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN PUMP.
Patient Age80 YR
Patient SexFemale
Patient Weight60 KG
-
-