RESPIRONICS, INC TRILOGY 202; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 1040000
Medical Device Problem Codes Mechanical Problem (1384); Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/09/2022
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging a ventilator would not power on.The device was not in patient use.The device has yet to be returned to the manufacturer for evaluation.At this time, we are unable to confirm the alleged malfunction.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Additional Manufacturer Narrative
The manufacturer previously reported a ventilator would not power on.Additional information received from the field service technician states the issue was related to the oxygen blending module board.No issue with power was noted.The device was returned to the customer unrepaired per the customer's request.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 202
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC
1001 murry ridge rd
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC
1001 murry ridge rd
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key14022393
Report Number2518422-2022-12653
Device Sequence Number2646491
Product Code CBK
UDI-Device Identifier00606959022096
UDI-Public00606959022096
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K093905
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1040000
Device Catalogue Number1040000
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/09/2022
Supplement Date Received by Manufacturer07/26/2022
Initial Report FDA Received Date04/06/2022
Supplement Report FDA Received Date07/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/03/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-