Model Number N/A
Medical Device Problem Code Positioning Failure (1158)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the poly would not seat per technique.The surgeon could not get the poly to seat so he opened a new one of the same size and it was implanted with no issues.There was no consequences or impact to the patient.No further event information available at the time of this report.
 
Additional Manufacturer Narrative
(b)(4).No product was returned or pictures provided; visual and dimensional evaluations could not be performed.The device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.D10: 42542006702 - tibia fixed cemented right size d stem extension use required - 64835319.42540000035 - all poly patella cemented 35 mm diameter - 65270980.42555803405 - tibial augment cemented half block right medial size cd 5 mm - 64353978.42545000510 - tibial central cone size x-small - 65034282.42555803205 - tibial augment cemented half block right lateral size cd 5 mm - 64333007.42560613513 - stem extension 6mm offset splined uncemented 13 mm diameter - 64739769.42560613514 - stem extension 6mm offset splined uncemented 14 mm diameter - 64876024.42556606605 - femoral distal augment cemented size 9, 9+ 5 mm thickness - 64614849.42556806605 - femoral posterior augment cemented size 9, 9+ 5 mm thickness - 65263702.42504606612 - femur cemented plus right size 9+ - 64287657.The following sections were updated: a2; b4; b5; d10; g3; g6; h1; h2; h10.This additional information does not change the previous investigation.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the returned device found signs of use (nicked and gouged) and the dovetail feature is flared and compressed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Damage to the articular surface can occur if the articular surface is not correctly placed and oriented before pushing it using the inserter instrument.Detailed instructions for inserting the articular surface are provided in persona the personalized knee surgical technique.The root cause is attributed to use error.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE FIXED BEARING CONSTRAINED POSTERIOR STABILIZED
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14022771
Report Number3007963827-2022-00084
Device Sequence Number2631919
Product Code MBH
UDI-Device Identifier00889024244474
UDI-Public(01)00889024244474(17)261103(10)65206363
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K123459
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42522600516
Device Lot Number65206363
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/22/2022
Supplement Date Received by Manufacturer04/08/2022
05/17/2022
Initial Report FDA Received Date04/06/2022
Supplement Report FDA Received Date04/13/2022
05/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE.
Patient SexFemale
-
-