| Model Number |
A343416A |
| Medical Device Problem Codes |
Component Missing (2306); Component Misassembled (4004)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/11/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the foley catheter tray had a piece missing.Stated that the missing piece was the portion to do the insertion of the catheter.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11:section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that the foley catheter tray had a piece missing.Stated that the missing piece was the portion to do the insertion of the catheter.
|
| |
|
Event or Problem Description
|
|
It was reported that the foley catheter tray had a piece missing.Stated that the missing piece was the portion to do the insertion of the catheter.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed manufacturing related.1 sample were confirmed to exhibit the reported failure.The device had not met specifications.The product was not used for patient treatment.The product caused the reported failure.Visual evaluation of the returned sample noted one opened (with original packaging), closed system foley tray.Visual inspection of the sample noted the all-silicone foley catheter missing from the tray upon return.This is out of specification which states, "assembly shall comply with the packaging layout".A potential root cause for this failure mode could be ¿incorrect operation¿.The device history record review could not be performed without a lot number.The investigation is concluded, and no additional action is required at this time.The device was returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: sterile unless package is opened or damaged, except for any individually packaged components within the tray which are not labeled as sterile.These components are not terminally sterilized.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Single use only.Do not resterilize.For urological use only.Warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage the catheter and may cause balloon to burst.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practices and applicable local, state and federal laws and regulations.Visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.Please consult product label and insert for any indications, contraindications, hazards, warnings, cautions and directions for use.Corrections: f,h.H11:section a through f the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the actual/suspected device was inspected.
|
| |
|