MAUDE Adverse Event Report: COOK INC GUARDIA ACCESSET EMBRYO TRANSFER CATHETER; MQF CATHETER, ASSISTED REPRODUCTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC GUARDIA ACCESSET EMBRYO TRANSFER CATHETER; MQF CATHETER, ASSISTED REPRODUCTION Back to Search Results
Model Number G53007
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/23/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Name and address - phone: (b)(6).Occupation ; technical manager/quality specialist.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
As reported, a hair was noted inside the packaging of a guardia accesset embryo transfer catheter.There was no patient involvement.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Event description: as reported, a hair was noted inside the packaging of a guardia accesset embryo transfer catheter.There was no patient involvement.Investigation ¿ evaluation.A visual inspection of the returned device was conducted.A document based investigation was also performed including a review of complaint history, device history record (dhr), the instructions for use (ifu), and quality control procedures.One, unopened guardia accesset embryo transfer catheter was returned for investigation.A black fiber was caught in the seal of the inner pouch which contained the guide catheter.A review of the device history record found no non-conformances related to the reported failure mode.A review of complaint history records shows no other complaints associated with the complaint device lot.Due to the individual nature of inspecting packaged product, one non conformance does not indicate additional non conformance`s in the lot.Additionally, there have been no other complaints filed for this lot.Therefore, there is no evidence to suggest that nonconforming product are in the field a review was of manufacturing procedures was performed, including quality control procedures.Controls were found to be in place, all packaged product of this type are visually inspected for particulate.Cook has concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: how supplied.Do not use the product if there is doubt as to where the product is sterile.Cook has concluded the cause of the complaint is manufacturing related.The employees who inspected the lot for foreign matter after packaging the product have completed defect awareness training.Per the quality engineering risk assessment, no further action is required.The appropriate personnel have been notified, and we will continue to monitor for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GUARDIA ACCESSET EMBRYO TRANSFER CATHETER
Common Device Name
MQF CATHETER, ASSISTED REPRODUCTION
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key14024178
Report Number1820334-2022-00537
Device Sequence Number19475420
Product Code MQF
UDI-Device Identifier00827002530072
UDI-Public(01)00827002530072(17)240928(10)14189294
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K173686
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date09/28/2024
Device Model NumberG53007
Device Catalogue NumberK-JETS-7019-ET
Device Lot Number14189294
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/25/2022
Supplement Date Received by Manufacturer05/18/2022
Initial Report FDA Received Date04/06/2022
Supplement Report FDA Received Date05/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/28/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-