• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH FEMORAL NECK SYSTEM PLATE 1 HOLE - STERILE; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH FEMORAL NECK SYSTEM PLATE 1 HOLE - STERILE; APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT Back to Search Results
Model Number 04.168.000S
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 01/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported on (b)(6) 2021, the patient underwent the open reduction internal fixation with the fns for a medial fracture of the femoral neck.The patient fell several weeks after the surgery and was under observation.Migration/cut-out was confirmed on (b)(6) 2022.The revision surgery for removal of the fns and the bipolar hip arthroplasty will be performed on (b)(6) 2022.Although the fracture was outside the scope of the osteosynthesis, the surgeon decided to perform the open reduction internal fixation with the femoral neck system (fns) to reduce bleeding because of the underlying disease.This event might have been caused by the fall.Revision surgery was done on (b)(6) 2022.Fns implants in question were removed and replaced with bha.The revision surgery was completed successfully with 1 hour delay.The delay in surgery is due to instrument breakage of other companies' products.No further information is available.This is report 1 of 4 for (b)(4).This report is for a femoral neck system plate.
 
Additional Manufacturer Narrative
If the information is unknown, not available or does not apply, the section/field of the form is left blank.Date of event: date of cut out is unknown date in 2022.Complainant part is not expected to be returned for manufacturer review/investigation.Device history report: part # 04.168.000s; lot # 312p097; manufacturing site: (b)(4); release to warehouse date: 09.August.2021; expiry date: 01.July.2031.A manufacturing record evaluation was performed for the finished device lot and no non-conformance was identified.Product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL NECK SYSTEM PLATE 1 HOLE - STERILE
Common Device Name
APPLIANCE, FIXATION, NAIL/BLADE/PLATE COMBINATION, MULTIPLE COMPONENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK GRENCHEN (CH)
solothurnstrasse 186
grenchen 2540
SZ   2540
Manufacturer Contact
kate karberg
1302 wrights lane east
west chester, PA 19380
8472871282
MDR Report Key14025264
Report Number8030965-2022-02204
Device Sequence Number20338399
Product Code KTT
UDI-Device Identifier10886982274557
UDI-Public(01)10886982274557
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K172872
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.168.000S
Device Catalogue Number04.168.000S-US
Device Lot Number312P097
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/08/2022
Initial Report FDA Received Date04/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ANTIROTSCR F/FEM NECK SYST F/CONSTRUCT L; BOLT F/FEM NECK SYST F/CONSTRUCT LENGTH; LOCKSCR Ø5 SELF-TAP L40 TAN
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-