• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL MICROBLENDER; MIXER, BREATHING GASES, ANESTHESIA INHALATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL MICROBLENDER; MIXER, BREATHING GASES, ANESTHESIA INHALATION Back to Search Results
Model Number MICBLNDR,W/O HOSES,DOM,H/L
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/11/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to vyaire medical that when the microblender is running at 50lpm or above on the auxiliary port and trying to get to 100% to check the tolerance, the blender starts alarming.The customer confirmed that there was no patient involvement associated with the reported event.
 
Additional Manufacturer Narrative
Vyaire medical file identification: (b)(4).The suspect device will not be returned for evaluation.Therefore, no root cause could be determined.However, technical support advised the customer to check the ball, it may be coming off the seat bypass and when he goes to the higher flows it is causing the unit to alarm.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROBLENDER
Common Device Name
MIXER, BREATHING GASES, ANESTHESIA INHALATION
Manufacturer (Section D)
VYAIRE MEDICAL
510 technology dr
irvine IL 92618
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs CA 92262
Manufacturer Contact
erika bonilla
510 technology dr
irvine, IL 92618
7149227837
MDR Report Key14026629
Report Number2021710-2022-15777
Device Sequence Number16802491
Product Code BZR
UDI-Device Identifier10846446002348
UDI-Public(01)10846446002348(11)19920101
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 03/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMICBLNDR,W/O HOSES,DOM,H/L
Device Catalogue Number03800A
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/11/2022
Initial Report FDA Received Date04/06/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/01/1992
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-