• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX AFX; BIFURCATED STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX AFX; BIFURCATED STENT GRAFT Back to Search Results
Model Number BA28-80/I20-40
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 02/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The devices involved in this event will be returned for evaluation.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx with duraply.
 
Event or Problem Description
The patient was initially treated for an abdominal aortic aneurysm (aaa) with implant of an afx bifurcated stent graft and an afx vela suprarenal proximal extension.Approximately three (3) years post initial procedure, the patient presented with a type ib endoleak.The physician elected to treat the patient by implanting a non-endologix prosthetic extension in the left leg sometime in year 2020.This type ib endoleak will be reported under a separate medical device report (mdr).Approximately two (2) years post secondary procedure, the patient presented with an endoleak (at indeterminate origin).The physician elected to convert patient to open repair and explant the afx devices.The final patient status was not reported.The explanted devices are available for return/evaluation.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation.A clinical evaluation of the adverse event/incident could not be completed.No medical records or medical imaging relevant to the adverse event/incident was received by endologix.Due to the absence of relevant medical records and reported medical imaging; device, use, procedure, and/or anatomy relatedness to this adverse event/incident could not be evaluated.The final patient status was not reported.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx with duraply.Corrections: h6 investigation type codes; remove 4118.H6 investigation finding codes; remove 3233.H6 investigation conclusion codes; remove 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key14026766
Report Number2031527-2022-00075
Device Sequence Number20367707
Product Code MIH
UDI-Device Identifier00818009012265
UDI-Public(01)00818009012265(17)190929
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Pharmacist
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 03/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/29/2019
Device Model NumberBA28-80/I20-40
Device Lot Number1595273-009
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/28/2022
Supplement Date Received by Manufacturer03/28/2022
Initial Report FDA Received Date04/06/2022
Supplement Report FDA Received Date06/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/14/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX VELA SUPRARENAL, SN (B)(4); NON-ENDOLOGIX PROSTHETIC EXTENSION
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-