Model Number 391.201
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that received a phone from or where surgeon was attempting to tension a 1.7mm using x1 391.201 lot#p279304.He was unable to pass the 1.7mm cable through the tensioner despite tensioner being all the way open.There is some sort of blockage in the tensioner.He called for another set and procedure was completed successfully with a 5 minute delay.There were no fragments generated.There was no adverse consequences.Other medical intervention required was unknown.This complaint involves one(1) device cable tensioner.This is report 1 of 1 for complaint pc-(b)(4).
 
Additional Manufacturer Narrative
Product complaint # pc-(b)(4).Complainant part is expected to be returned for manufacturer review/investigation but has yet to be received.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h4, h6: part # 391.201.Synthes lot # p279304.Supplier lot # n/a.Release to warehouse date: july 31, 2017.Supplier:(b)(4).No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.The product was returned to depuy synthes for evaluation.The depuy synthes team forwarded the device to resolve surgical which conducted a visual inspection of the returned device.Visual analysis of the returned sample was performed on cable tensioner.Through a visual investigation it was found that device had damaged cable/cable fragments or debris blocking the opening in the collet tip.This blockage was not able to be cleared from device, therefore the device in question is not able to be tested.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for the cable tensioner.The received condition of the cable tensioner confirmed that damaged cable/cable fragments or debris blocking the opening in the collet tip there is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D9.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CABLE TENSIONER
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
RESOLVE SURGICAL TECHNOLOGIES
375 river park circle
marquette MI 49855
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key14027197
Report Number1833824-2022-00029
Device Sequence Number14111359
Product Code LXH
UDI-Device Identifier10886982198204
UDI-Public(01)10886982198204
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number391.201
Device Catalogue Number391.201
Device Lot NumberP279304
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/10/2022
Supplement Date Received by Manufacturer05/05/2022
Initial Report FDA Received Date04/06/2022
Supplement Report FDA Received Date05/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/31/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-