Model Number Z9002
Medical Device Problem Codes Inaccurate Delivery (2339); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/16/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If implanted; give date: not applicable as the iol was not implanted.If explanted; give date: not applicable as the iol was not explanted.The device has not returned for evaluation.Therefore, a visual analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain the missing information.However, to date it has not been received.All pertinent information available to johnson & johnson surgical vision, inc., has been submitted.
 
Event or Problem Description
Received report from health care provided (hcp) indicating the intraocular lens (iol) tore when the surgeon pushed it through the injector.The customer explained that both the iol and the cartridge touched the eye.There was no capsule tear, no vitrectomy or sutures needed.And no medication outside the regular standard of care.The iol partially went into the patient¿s eye.The doctor then removed the cartridge and iol.A new cartridge and iol were given to the scrub nurse.The procedure was completed with the backup lens without any complications.The explanted iol will not be returned as it was discarded by the customer.No further information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS CL IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key14027328
Report Number3012236936-2022-00794
Device Sequence Number6045441
Product Code HQL
UDI-Device Identifier05050474530744
UDI-Public(01)05050474530744(17)260625
Combination Product (Y/N)N
Initial Reporter StateAR
Initial Reporter CountryUS
PMA/510(K) Number
P880081
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/25/2026
Device Model NumberZ9002
Device Catalogue NumberZ900200225
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/16/2022
Initial Report FDA Received Date04/06/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/25/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-