BAUSCH + LOMB, INC STELLARIS MICRO INCISION VACUUM PACK; UNIT, PHACOFRAGMENTATION Back to Search Results
Model Number BL3118
Medical Device Problem Code Particulates (1451)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing and sterilization records were reviewed and found to be acceptable.The product was not returned so we are unable to investigate for the root cause.The lot history, trend analysis, risk analysis and/or directions for use review were considered acceptable, with the product performing within anticipated rates.No corrective action is required.If the product is returned at a later date, this investigation will be re-opened.At this time, the investigation is considered complete.
 
Event or Problem Description
A user facility in (b)(6) reports that during surgery, the surgeon found a microscopic transparent filament in the patient¿s eye.The surgeon asserts that the transparent filament came from the sterile medical device.The filament was extracted.The patient outcome is unknown.
 
Additional Manufacturer Narrative
One small plastic vial/specimen cup was returned in a plastic bag.None of the pack components were received.The part number and lot number of the pack cannot be determined.The vial contains an unknown fluid and a light colored, transparent, long, particle can be seen floating in the fluid inside of the vial.The particle is tube-like in shape, and is similar to a straw in that it appears to be hollow in the middle.The particle is flexible and easily damaged.The bend in the straw-like particle was caused from the forceps that were used to retrieve it from the fluid.The particle is approximately 1119 microns long by 21 microns wide.This particle was sent to an external laboratory for analysis.The analysis of the particle indicated it consisted primarily of polyamide 6 (nylon 6) with lesser concentrations of titanium dioxide pigment.The particle sample was microscopically examined, photographed, and measured using a camera-equipped optical stereo microscope.The sample was analyzed using an energy-dispersive spectrometer (eds) equipped scanning electron microscope(sem) and a fourier transform infrared (ftir) spectrometer.Eds analysis indicated that the particle consists primarily of carbon and oxygen.Lesser concentrations of titanium were also detected.This is consistent with organic material and titanium dioxide pigment.The ftir analysis of the particle produced a spectrum consistent with that of polyamide 6 (pa6).Laboratory analysis indicates the particle consists primarily of polyamide 6 (nylon 6) with lesser concentrations of titanium dioxide pigment.These materials are not used in the bl3118 product.The source of the returned particle cannot be determined.Therefore the root cause is unknown/inconclusive.It was reported that there were no clinical consequences.The investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STELLARIS MICRO INCISION VACUUM PACK
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
BAUSCH + LOMB, INC
1400 n goodman st
rochester NY 14609
Manufacturer (Section G)
BAUSCH + LOMB, INC
3365 tree court industrial blv
st. louis MO 63122
Manufacturer Contact
juli moore
3365 tree court industrial blvd
st. louis, MO 63122
MDR Report Key14027995
Report Number0001920664-2022-00046
Device Sequence Number2570093
Product Code HQC
UDI-Device Identifier10757770057389
UDI-Public(01)10757770057389(17)261016
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K063331
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBL3118
Device Catalogue NumberBL3118
Device Lot Number222
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/10/2022
Supplement Date Received by Manufacturer04/13/2022
Initial Report FDA Received Date04/06/2022
Supplement Report FDA Received Date05/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/29/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-