• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG PROTEGO T 65; ICD LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG PROTEGO T 65; ICD LEAD Back to Search Results
Model Number 379967
Medical Device Problem Codes Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 03/29/2022
Type of Reportable Event Serious Injury
Event or Problem Description
After an implantation period of approx.80 months, oversensing with inappropriate therapies was reported.The physicians decided to switch off the tachy therapies.The lead remains implanted.No adverse patient events were reported.Should additional information be received, this file will be updated.
 
Additional Manufacturer Narrative
The icd and the lead under complaint were not returned for analysis.The analysis is therefore based on the inspection of the quality documents associated with the manufacture of these particular devices as well as on the returned data.The manufacturing processes for the lead and the icd were re-investigated and all production steps were performed accordingly.There was no sign of any inconsistency during the manufacturing process.Particularly the final acceptance tests proved the devices functions to be as specified.The returned data have been analyzed.The analysis of the available iegms confirmed the presence of noise in the right ventricular channel.Besides that, the data showed no anomalies.Based on the available information the root cause of the noise could not be conclusively determined.However, the integrity of the lead cannot be assured.There was no indication of an icd malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTEGO T 65
Common Device Name
ICD LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key14029075
Report Number1028232-2022-01760
Device Sequence Number4776050
Product Code NVY
UDI-Device Identifier04035479121686
UDI-Public04035479121686
Combination Product (Y/N)N
PMA/510(K) Number
P980023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2017
Device Model Number379967
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/06/2022
Supplement Date Received by Manufacturer06/27/2022
Initial Report FDA Received Date04/06/2022
Supplement Report FDA Received Date06/28/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/25/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient SexUnknown
-
-